TE-20-800 Ambient air sampler, particulate sizing instrument, Ambient non-Viable Cascade Sampler Impactor |
 |
 |
| Support |
 |
|
|
|
|
|
|
Eight Stage Ambient Sampler System Includes: 8 Aluminum Orifice Stages, 8 Stainless Steel Collection Plates, Back up 81mm Filter Holder. 11 Silicone O-Ring Gaskets, Vacuum Pump Assembly and Carrying Case complete |
| Features: |
High-Capacity Pre-impactor ensures accurate separation of
large particles.
High
Sample Flow-Rate of 28.3 ALPM
Size
separation from >9nm to <0.3nm
Optional
volumetric flow-controlled vacuum system
Corrosion-resistant
alloy aluminum with silicone seal |
| AERODYNAMIC
SIZING AND DISTRIBUTION OF AMBIENT PARTICULATE WITH UNPRECEDENTED
PRECISION...COST... AND EASE OF OPERATION |
Time
proven virtual elimination of particle bounce and re-entrainment.
Highest
proven collection efficiency, reproducibility and accuracy
Aerodynamic
particle separation in nine (9) size fractions
81mm
Sample Substrates for easy handling and weighing with analytical
balance.
Incorporates
U.S. Federal standards for the Ideal Cascade Impactor"
And meets OSHA, EPA and ACGIH design requirements for respirable/non
respirable fractionation. |
Cut-points
for the Tisch Environmental Eight Stage Non-Viable Impactor
at 28.3 LPM |
STAGE
|
MICROMETER
CUT-POINT OPERATING AT 28.3 lpm |
| Pre-Seperator |
10.0
and above |
| 0 |
9.0 |
| 1 |
5.8 |
| 2 |
4.7 |
| 3 |
3.3 |
| 4 |
2.1 |
| 5 |
1.1 |
| 6 |
.7 |
| 7 |
.4 |
| Shipping
Info. |
Size:
8" x 23" x 13" Weight: 21 lbs. |
|